Model |
HR-1100 |
|
Display |
LCD |
|
Waterproof specifications |
Equivalent to IPX5 |
|
Input connector |
ASP |
|
Input type |
Thermocouple Input: Type E, K |
|
Channel |
1ch |
|
Signal source resistance |
MAX.1k ohm |
|
Measurement Range |
1°C Resolution |
E |
-200 to 800°C |
|
K |
-200 to 1370°C |
|
0.1°C Resolution |
E |
-104.9 to 504.9°C
(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
|
|
K |
-104.9 to 504.9°C
(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
|
|
Measurement Accuracy |
1°C Resolution |
0°C to |
± (0.1% of reading +1°C) |
|
to 0°C |
± (0.5% of reading +1°C) |
|
0.1°C Resolution |
0°C to |
± (0.05% of reading + 0.2°C) |
|
to 0°C |
± (0.15% of reading + 0.2°C) |
|
Reference junction compensation accuracy |
±0.2°C at 25°C±10°C |
|
Temperature Coefficient*1 |
±0.02×Δt°C (Add the value obtained by multiplying the excess temp.Δt[°C] by a coefficient to the total accuracy. e.g. ±0.3°C is added at room temp. of 0°C or 50°C)
|
|
Operation Environment |
0 to 50°C, 0 to 80%RH (Non condensing)
|
|
Storage Environment |
-20 to 50°C, 0 to 85%RH (Non condensing)
|
|
Battery life |
approx.900h |
|
Power supply(Dry battery) |
4 AA Alkaline batteries |
|
Power supply(AC power) |
- |
|
Sampling rate |
approx.200ms |
|
Linearlizer |
Digital Linearizer method (Compliant with JIS C1602-2015) |
|
Dimensions(mm) |
approx. 82(w)×166(H)×36(D) [Protruded parts excluded] |
|
Weight |
approx. 350g [Include Batteries] |
|
Supplied accessories |
Manual,Test report,Soft case, |
|
Hand strap,4 AA Alkaline batteries |
|
Compatible standards |
EMC:EN61326-1:2013, EN 61326-2-1:2013 classA Table2( Industrial)
RoHS:IEC EN 63000:2018
|