Thông số kỹ thuật
Disadvantages are obvious for the hardness measurement commonly used like Brinell, Rockwell, Vickers, Leeb, etc.
* Rockwell and Brinell with heavy loading force and big indentation, lead to serious destruction on sample surface.
* Vickers apply optical measurement, but only professional technicians can operate smoothly, impossible to measure hardness of heavy work piece, installed machinery and permanently assembled parts.
* Leeb hardness tester apply rebound and indirect method to measure hardness, easily lead to big deviation when convert to Brinell, Rockwell and Vickers scales.
* Ultrasonic hardness tester HUH-5M applies ultrasonic contact impedance method to do comparative hardness measurement for testing pieces, with advantages of high accuracy, efficiency, portable and easy operation.
* Ultrasonic hardness tester HUH-5M is widely used to measure hardness of small forgings, cast material, weld inspection, heat affected zone, Ion-nitride stamping dies and molds, forms, presses, thin walled parts, bearings, tooth flanks, etc.
Manual Probe UCI-1K/UCI-2K/UCI-5K/UCI-10K
Manual Probes Features:
* Introduced latest ultrasonic sensor technology
* Obtain accurate hardness value without auxiliary devices
* Support 360° quick measuring
* Excellent human engineering design easy to hold
* Suitable dimension and weight of probe is convenient to control
* Quick and accurate measuring for edge and fixed position of specimen
* Micro indentation, Non- Destructive for specimen
Specifications:
Product Name |
UCI Ultrasonic Hardness Tester |
Model |
HUH-5E |
Loading Force |
10N Motorized Probe |
Indenter |
136°Vickers Diamond Indenter |
Measuring Range |
HB: 85-650;HV 80-1599; HRC 20-70;HRB: 41-100; HRA: 61-85.6
HS: 34.2-97.3;Mpa: 255-2180N/mm
|
Measuring Accuracy |
HV:±3%HV; HRC:±1.5HRC; HB:±3%HB; HLD:±5HLD; |
Measuring Direction |
Support 360° |
Data Storage |
To save 1000-groups of measuring data and 20-groups of calibration data |
Hardness Scale |
HV,HB,HRC, HLD |
Data Display |
Loading force, Testing-times, Testing result, Average,
Maximum,Minimum,Deviation and Conversion scale. |
Hardness Indication |
LCD display |
OperatingEnvironment |
Temperature:-10℃~50℃; Humidity: 30%~80%R.H |
Operating Voltage |
DC 6V |
Instrument Dimensions |
160x80x31mm |
Net Weight |
Approximate 500g(Without probe) |
Standard Delivery:
Standard Delivery
|
quantity
|
Standard Delivery
|
quantity
|
Standard Delivery
|
quantity
|
Standard Delivery
|
quantity
|
HUH-5E main unit
|
1
|
Probe Cable
|
1
|
Anti Vibration Case
|
1
|
English Operation Manual
|
1
|
UCI-1000 Motorized Probe
|
1
|
Com Cable
|
1
|
Screw
|
1
|
Warranty Card
|
1
|
Standard Rockwell Block
|
1
|
USB Cable
|
1
|
Qualification Certificate
|
1
|
Packing List
|
1
|
Manual Probe Specifications:
Probe Type
|
UCI-1K
|
UCI-2K
|
UCI-5K
|
UCI-10K
|
Testing force
|
10N
|
20N
|
50N
|
98N
|
Diamter
|
22mm
|
22mm
|
22mm
|
22mm
|
Length
|
154mm
|
154mm
|
154mm
|
154mm
|
Oscillating Rod Diameter
|
2.4mm
|
2.4mm
|
3mm
|
3mm
|
Roughness of measuring surface
|
Ra<3.2um
|
Ra<5um
|
Ra<10um
|
Ra<15um
|
Min weight of test material
|
0.3kg
|
0.3kg
|
0.3kg
|
0.3kg
|
Minimum thickness
|
2mm
|
2mm
|
2mm
|
2mm
|
Motorized Probe Specifications:
Probe Type
|
UCI-300
|
UCI-500
|
UCI-800
|
UCI-1000
|
Testing force
|
3N
|
5N
|
8N
|
10N
|
Diamter
|
46mm
|
46mm
|
46mm
|
46mm
|
Length
|
197.5mm
|
197.5mm
|
197.5mm
|
197.5mm
|
Oscillating Rod Diameter
|
3.7mm
|
3.7mm
|
3.7mm
|
3.7mm
|
Min weight of test material
|
0.3kg
|
0.3kg
|
0.3kg
|
0.3kg
|
Minimum thickness
|
2mm
|
2mm
|
2mm
|
2mm
|
Testing Stand HU-100: